Week 3
This week I saw my first return client. She came in over the summer for a complete hearing evaluation and at that point, decided that she did not want hearing aids. However, after thinking it over, she changed her mind and came back in to get fitted. Before discussing which hearing aids to use, I found her MCL (most comfortable level) and UCL (uncomfortable level).
Martin defines MCL as the level at which speech is most comfortable to the client. Several studies have been done to determine test-retest reliability of MCL. In one particular study, it was shown that about 95% of MCLs vary within 8 to 12 dB from when it was first obtained. Ventry and Johnson in 1978 found in their study of 100 hearing impaired listeners (all had sensorineural hearing loss) that MCLs obtained from a descending method were 20-23 dB higher than the ascending results.
UCL is defined as the level where speech becomes uncomfortable or a person to listen to for an extended period of time. Normal listeners typically have a UCL around 100 dB (although I think mine is even less than that). UCL is often obtained in order to set the OSPL 90 of the hearing aid. Typically, test-retest reliability of UCL is 4 to 8 dB. Some even believe acoustic reflex threshold is a good prediction of UCL but this is a controversial issue.
It is recommended that ASHA develops a standardization in ways to obtain both UCL and MCL because of the wide variability.
Source: Punch, J., Joseph, A., & Rakerd, B. American Journal of Audiology, 13, 144-157.
Martin defines MCL as the level at which speech is most comfortable to the client. Several studies have been done to determine test-retest reliability of MCL. In one particular study, it was shown that about 95% of MCLs vary within 8 to 12 dB from when it was first obtained. Ventry and Johnson in 1978 found in their study of 100 hearing impaired listeners (all had sensorineural hearing loss) that MCLs obtained from a descending method were 20-23 dB higher than the ascending results.
UCL is defined as the level where speech becomes uncomfortable or a person to listen to for an extended period of time. Normal listeners typically have a UCL around 100 dB (although I think mine is even less than that). UCL is often obtained in order to set the OSPL 90 of the hearing aid. Typically, test-retest reliability of UCL is 4 to 8 dB. Some even believe acoustic reflex threshold is a good prediction of UCL but this is a controversial issue.
It is recommended that ASHA develops a standardization in ways to obtain both UCL and MCL because of the wide variability.
Source: Punch, J., Joseph, A., & Rakerd, B. American Journal of Audiology, 13, 144-157.